HomeAllPetroFLAG Analyzer System for TPHin Soil
soil icon

Soil

PetroFLAG Analyzer System for TPH in Soil

Commercially available kit for monitoring TPH in soil.

Customer link and technical handbooks

Category

Commercial kits

Environmental matrix

Soil

Parameters analysed

TPH (aromatic and aliphatic fractions, C8 – C44)

Detection limits

The lower limit of detection (LLD) for most hydrocarbons is 20 ppm, except for aged gasoline which has an LLD of 1,000 ppm. Using the standard 10- gram sample, the analyser has the ability to quantify all hydrocarbons up to 2,000 ppm.

Sensor functionalities

Turbidimetric detection.

Costs

~ 90 € (10 samples)

Advantages

  • Requires moderate skills to operate.
  • Appropriate for field measurements.
  • Responds to a broad range of petroleum products.
  • Results available within minutes.

Limitations

  • Qualitative or semiquantitative method.
  • Gasoline hydrocarbons (< C8) are not detected.
  • Possible interference with natural-occurring hydrocarbons and soil moisture.
  • Some prior knowledge of the contaminant required.

Other technical specifications

Results available in 15 min. Different response factors are applied depending on the contaminants.
Calibration required.